Description
Precision Measurements: The QuadTech 1865
provides resistance measurement capability from 1kΩ to
100TΩ (test voltage dependent) with a basic accuracy of
0.5%. To meet test requirements for a wide range of
devices, the test voltage is fully programmable up to
1000V DC. Sensing the proper resistance measurement range
is done automatically, thus eliminating setup errors; and
in order to eliminate lead or fixture errors, the operator
can easily initiate an automatic zeroing routine of test
leads.
Pass/Fail Testing: Measured results are
automatically compared to an operator programmed limit for
pass/fail testing. Pass/Fail indicator lights or a pass/fail
output from the instrument's I/O interface provide a clear
indication of the measured results.
Program and Data Storage: Test setup
conditions and measured results can be stored in instrument
memory or on 3 1/2" disk when using the extended program
storage option.
Automated Testing: For automated system
applications, the instrument includes an I/O interface
connection with remote start and pass/fail outputs. An
optional IEEE-488 interface is also available which enables
the 1865 to be used under computer control.
Current Display Mode: Besides a readout
of insulation resistance in ohms, the operator can select a
display mode for reading the measured current directly.
Safety Features: The 1865 provides
safety features such as current limiting, a warning
indicator when high voltage is active, and safety interlock,
all for protection of the operator.
Component Test Fixture: An accessory
fixture is available for use with the 1865 which accomodates
a variety of component types, including radial, axial and
chip components. It's shielded case reduces electrical noise
and interference and includes a cover interlock switch and
remote start for maximum operator safety.
Why Buy?
- For precision high resistance measurements over a wide
voltage range
- For automated testing requiring remote start and
pass/fail output
- To data log test results via RS-232 or IEEE-488
interface
- To transport test results to an Excel spread sheet via
3 ½” disk
Features
- Measurements from 1kΩ to 100TΩ, with Auto Ranging
- Direct Reading of Measured Resistance or Current
- 0.5% Basic Measurement Accuracy
- Programmable Test Voltage from 1 to 1000VDC
- Large LCD Display for Menu Programming and Test
Results
- Fully Programmable Measure Cycle Times
- Averaging for Increased Accuracy
- Storage of Multiple Test Setups for Quick Startup
- RS-232 and PLC Interfaces Standard, IEEE-488 Optional
- Keypad Lockout with Password Protection
- Automatic Test Lead Zeroing To Eliminate Connection
Errors
- Limit Entry for Pass/Fail Testing
- Safety Interlock and High Voltage Indicator for
Operator Safety
Product Package
The following accessories come standard with the unit:
- 150073 1865 Instruction Manual
- 4200-0300 AC Power Cable
- 800014/15 Capacitor Adapters (100kΩ & 1MΩ)
- 520049 Power Line Fuse (2.5A, 250V)
- 630019 Interlock Connector
- Calibration Certificate Traceable to NIST
|