SRX1 Series NAVAIR Resistance Standard

Double click on above image to view full picture

Zoom Out
Zoom In

More Views

Photo represents example of series. See datasheet for more details on any specific model.

Resources

SRX1 Series NAVAIR Resistance Standard

This is a series made up of a number of models. To find your specific model either click SHOW ALL Button or use the Filters Boxes, below. Then select the model by clicking the model number (blue text in table).

  • Range from 1 Ω - 1 MΩ
  • Stabilities as good as 5 ppm
  • Minimal Change from power cycling
  • The US Navy's Resistance Standard


The SRX1 Series are extremely stable, precise, laboratory or portable resistance standards.

The SRX1 Series resistance standards are ideal for any applications inside or outside of a laboratory environment within the temperature range of 18°C to 28°C. They are rugged and small, and they have virtually zero temperature coefficient. Each unit has a temperature chart showing its accuracy range in 0.5°C increments. The unit requires no oil or air bath because of the low temperature coefficient.

SPECIFICATIONS

Model SRX1- Nominal Value (Ω) Adjustment to Nominal (ppm) Stability 1 year (ppm) Max Voltage(V)
1 1 <20 <10 <0.32
10 10 <20 <10 <1
100 100 <15 <7 <2
1k 1 k <10 <7 <10
10k 10 k <15 <5 <30
100k 100 k <15 <7 <100
1M 1 M <22 <10 <100

Test Conditions: Four-terminal Kelvin measurements, low power, at 23°C; two-terminal for 1 MΩ. Initial calibration data, traceable to SI, is provided.
Terminals:The binding posts are tellurium copper for low thermal emf and low resistance. A case ground terminal is also provided. 100 kΩ and under: Four 5-way binding posts for 4-terminal measurement.1 MΩ and over: Two 5-way binding posts for 2-terminal measurement.
Operating Conditions: 18-28°C; <80% RH.
Maximum Applied Power: As indicated in table for Max. Voltage above. Negligible change in resistance value will result from applying voltages up to the maximum voltage as indicated in the table above. Application of up to 5 W is allowed; change in resistance value from the max voltage up to a total of 5 W will be <300 ppm.
Change Resulting from Power Cycling: Value of the standard will remain within specifications after unlimited applications of power cycles of up to 5 W.
Resistance Change with Temperature: Outside the range of +18°C to +28°C, the Standard Resistors shall not change more than 3 ppm from their reported calibration value.
Change Resulting from Temperature Cycling: The standard resitors shall have a temperature cycling change of 2 ppm or less for a cycle from 23° to 0°C to 23°C and for a temperature cycle from 23°C to 40°C to 23°C.
Non-Operating Storage Conditions: 18-28°C; RH not controlled.
Dimensions: 8.6 cm H x 10.2 cm W x 12.7 cm D (3.4" x 4.0" x 5"). Weight: 1 lb, nominal