The 1864-9700 is an analog high-resistance megohmmeter for insulation-resistance and leakage measurements on wire, cable, capacitors, components and devices when the applied DC test voltage must match the requirement under test. Its selectable source and direct analog indication let a laboratory or inspection bench test at more than a small set of fixed voltage points.
- Test voltage: 200 test voltages, from 10 Vdc to 109 Vdc in 1 V steps and from 100 Vdc to 1090 Vdc in 10 V steps.
- Resistance measurement: 50 kΩ to 200 TΩ, with useful range dependent on the selected voltage band and meter position; the useful range reaches 200 TΩ at 500 Vdc to 1090 Vdc, while lower-voltage bands have useful limits of 2 TΩ or 20 TΩ.
- Voltage accuracy across the unknown: ± 2%; short-circuit current: approximately 5 mA.
- Guard and Ground terminals permit measurements of grounded or ungrounded two- or three-terminal devices.
- Resistance accuracy has a minimum reading of 0.5; on ranges 1 through 5 it is ±2(meter reading+1)%—±3% at a meter reading of 0.5.
Stated accuracy applies at 0 °C to + 40 °C and < 70% RH.
Set the insulation or leakage test voltage your requirement calls for
The 1864 Megohmmeter is for high-resistance, insulation-resistance and leakage measurements where the applied test voltage is part of the acceptance requirement. It is suited to wire and cable insulation, capacitor insulation, and a wide variety of components and devices, with a stated measurement span from 50 kΩ to 200 TΩ.
Its defining advantage is voltage selection. The 1864 provides 200 test voltages: any setting from 10 Vdc to 109 Vdc in 1 V steps, followed by 100 Vdc to 1090 Vdc in 10 V steps. This lets an engineer set the DC test condition required for the device instead of treating every insulation system as though it were specified at the same voltage. The selectable range includes common capacitor test voltages.
Confirm the required resistance range at the intended voltage, not only the overall 50 kΩ to 200 TΩ span. At 10 Vdc to 50 Vdc, the useful range at 2.5% of scale is 2 TΩ. At 50 Vdc to 100 Vdc and 100 Vdc to 500 Vdc, it is 20 TΩ; at 500 Vdc to 1090 Vdc, it is 200 TΩ. The lowest readable and useful values also change with voltage band, so the required operating point should be part of the quotation request.
Read high resistance directly at the bench
The front-panel analog meter provides direct-reading indication, and the lighted range switch identifies the multiplier for the selected range and voltage. This gives the operator the active test condition and resistance indication at the instrument. An analog output is provided where remote indication is required.
Resistance accuracy is expressed in terms of meter reading and has a minimum reading of 0.5. On ranges 1 through 5, accuracy is ±2(meter reading+1)%; the stated example at a meter reading of 0.5 is ±3%. Ranges 6 through 8 add the range terms listed in the specifications. This distinction is important when the required measurement is near the low end of a scale rather than at a higher meter reading.
Stable power supplies and a feedback voltmeter circuit minimize drift and adjustment during use. Voltage accuracy across the unknown is ± 2%, and short-circuit current is approximately 5 mA.
Use guard connections when the setup has unwanted leakage paths
The front panel has (+) Unknown, (-) Unknown, Guard and Ground terminals, permitting measurements of grounded or ungrounded two- or three-terminal devices. For a high-resistance measurement, leakage over cable insulation, fixture surfaces or contaminated insulating material can influence the observed result. Where the test object or fixture requires a three-terminal arrangement, the Guard connection allows the applicable guarded measurement connection described in the manual.
The specified accessories are the 630018 Lead Set, 630018/S Shielded Lead Set and 1863-11 Resistivity Test Fixture. The 1863 and 1864 Megohmmeters User and Service Manual provides connection and application guidance for insulation testing, capacitor insulation resistance, guarded three-terminal measurements, remote shielded measurements and resistivity measurements.
High-voltage and environmental considerations
A caution high-voltage indicator warns when voltage is present at the terminals. The manual identifies voltages above 45 V as hazardous. Use maximum insulation, minimize bare conductors, prevent accidental contact with live components and keep unrequired personnel away when hazardous voltages are used.
Stated accuracy applies from 0 °C to + 40 °C at < 70% RH. This condition matters in high-resistance work because humidity, contamination, lead insulation and fixture condition can create unintended leakage paths. For capacitor insulation tests, consult the manual for its charging-time, discharge-time and large-capacitor application guidance before establishing a test method.
The IET difference
The 1864 is manufactured using the same design and construction as the original GenRad megohmmeters. Its portable flip-tilt case protects the instrument during transit and storage and serves as a stand when the meter is in use. Model 1864-9700 includes an Instruction Manual and a Calibration Certificate Traceable to SI.
The warranty covers defects in material and workmanship for one year after original shipment. When properly used, an instrument found not to perform in accordance with applicable IET specifications during that period will be repaired or, at IET's option, replaced at no charge when returned to IET.
Specify the measurement, connection and accessories
Request a quote for 1864-9700 and provide the required DC test voltage, expected resistance range and required meter-scale position. Also identify whether the device is grounded or requires a guarded three-terminal connection, and whether you need the 630018 Lead Set, 630018/S Shielded Lead Set or 1863-11 Resistivity Test Fixture. For complete operating, safety, service and calibration information, consult the user and service manual and the 1863 and 1864 Megohmmeter datasheet.
| Voltage setting | Resistance range and useful ranges |
|---|---|
| 10 Vdc to 50 Vdc | Rmin (full scale left end, 0.5 rdg.): 50 kΩ; Rmax (right end, 10% of scale): 500 GΩ; useful range (2.5% of scale): 2 TΩ; useful ranges: 7 (recommended limit). |
| 50 Vdc to 100 Vdc | Rmin (full scale left end, 0.5 rdg.): 200 kΩ; Rmax (right end, 10% of scale): 5 TΩ; useful range (2.5% of scale): 20 TΩ; useful ranges: 8. |
| 100 Vdc to 500 Vdc | Rmin (full scale left end, 0.5 rdg.): 500 kΩ; Rmax (right end, 10% of scale): 5 TΩ; useful range (2.5% of scale): 20 TΩ; useful ranges: 7 (recommended limit). |
| 500 Vdc to 1090 Vdc | Rmin (full scale left end, 0.5 rdg.): 5 MΩ; Rmax (right end, 10% of scale): 50 TΩ; useful range (2.5% of scale): 200 TΩ; useful ranges: 8. |
| Specification | Value |
|---|---|
| Resistance accuracy | Minimum reading is 0.5. Range 1-5: ±2(meter reading+1)%. Example: if meter reading is 0.5, accuracy is ±2(0.5+1)% = ±3%. Range 6: ±(2(meter reading+1)%+2%). Range 7: ±(2(meter reading+1)%+3%). Range 8: ±(2(meter reading+1)%+5%). |
| Test voltages | 200 test voltages: 10 Vdc to 1090 Vdc. 10 Vdc to 109 Vdc in 1 V steps; 100 Vdc to 1090 Vdc in 10 V steps. |
| Voltage accuracy | ± 2% across unknown. |
| Short-circuit current | Approximately 5 mA. |
| Meter display | Analog meter; caution high-voltage indicator. |
| Analog output | Analog output. |
| Input terminals | Front-panel mounted: (+) Unknown (Red), (-) Unknown (Red), Guard (Red), Ground (Gold). |
| Specification | Value |
|---|---|
| Power | 100 - 125 or 200 - 250 V, switch selectable; 50 - 400 Hz; 13 W. |
| Fuse | For 90 to 240 V operation: T 250mA, 250 Vac, 5 x 20 mm fuse. |
| Operating environment | 0 °C to + 40 °C, stated accuracy, < 70% RH. |
| Storage environment | -20 °C to + 60 °C, < 80% RH. |
| Dimensions | 16.8 x 25.4 x 17.1 cm (6.63 x 10 x 6.75 in.). |
| Weight | 4.4 kg (9.5 lbs) net; 7.0 kg (14 lbs) shipping. |
| Item | Description |
|---|---|
| Model | 1864-9700 1864 Megohmmeter. |
| Included documentation | Instruction Manual; Calibration Certificate Traceable to SI. |
| Sold separately | 630018 Lead Set; 630018/S Shielded Lead Set; 1863-11 Resistivity Test Fixture. |