HRRS Series High Resistance Decade Box

Double click on above image to view full picture

Zoom Out
Zoom In

More Views

Photo represents example of series. See datasheet for more details on any specific model.

Resources

HRRS Series High Resistance Decade Box

To find the right model for you, please select:
(After selection, click the blue text below to pull up a model)

  • Resistance from 10 Ω to 100 GΩ
  • High-accuracy: up to .01%, 3 Accuracy Classes Available
  • Voltage up to 2000 V
  • Excellent TC: as low as 5 ppm/°C
  • Low voltage coefficient: as low as 0.2 ppm/V
  • High-performance solid silver-alloy switches
  • Rack mounting available



The High Resistance (HRRS Series) is an economical, high-performance, high-resistance decade box for all laboratory, test, and calibration needs. The High-resistance design is implemented with special resistors, switches, binding posts, and construction-design to allow high resistance performance without sacrificing other electrical properties. These instruments are useful wherever small currents and high resistances are required,

The HRRS Series employs state-of-the-art precision resistors of various types for high accuracy, high stability, and low temperature and voltage coefficients. The standard models offer a choice of one to seven decades. For custom units, models up to 11 decades are available . For higher voltage or higher resistance, models up to >10 kV and 10 TΩ are also available in the HRRS-5kV & -10kV Series.

The binding posts are for the HRRS high resistance decade boxes are insulated with Kel-F high-resistance, non-moisture-absorbing material ensure low leakage.
With a resolution as low as 10 Ω and a maximum available resistance of over 100 G&Omega, the HRRS Series may be used for exacting precision high-value resistance applications.

The HRRS series decade box complements the HARS series which provides resistance steps as low as 1 mΩ. The units may be rack-mounted to serve as components in measurement and control systems.

HRRS High Resistance Decade Box Applications

  • calibration of meters and megohmmeters,
  • checking of electrochemical and biomedical sensors and instruments.
  • testing high-impedance amplifiers and the insulation of low-power circuits.

HRRS Basic Specifications

Resistance per step
Total decade resistance
Accuracy
Max voltage per step (V)
Maximum Voltage (V)
Temperature coefficient (ppm/C)
Voltage coefficient (ppm/V)
Stability (ppm/yr)
Resistor Type
Q
B
F
10Ω 100Ω 0.01% 0.03% 0.10% 2.5 25 15 0 10 Wirewound, non-inductive
100Ω 1kΩ 0.01% 0.03% 0.10% 8 80 5 0 10 Wirewound, non-inductive
1kΩ 10kΩ 0.01% 0.03% 0.10% 23 230 5 0 10 Wirewound, non-inductive
10kΩ 100kΩ 0.01% 0.03% 0.10% 70 700 5 0 10 Wirewound, non-inductive
100kΩ 1MΩ 0.01% 0.03% 0.10% 230 2000 5 0 10 Wirewound, non-inductive
1MΩ 10MΩ 0.01% 0.03% 0.10% 1000* 2000 15 0.2 25 Metal-oxide film
10MΩ 100MΩ 0.03% 0.10% 1% 1000* 2000 15 0.2 50 Metal-oxide film
100MΩ 1GΩ 0.10% 0.20% 1% 1000* 2000 15 1.5 100 Metal-oxide film
1GΩ 10GΩ 0.20% 0.50% 1% 1000* 2000 50 5 500 High-voltage film
10GΩ 100GΩ 0.50% 1% 1% 1000* 2000 50 5 500 High-voltage film
100GΩ 1TΩ See HRRS-5kV Series for this model.
*To apply up to 2000 V for the first step, use the preceding decade at its "10" position. e.g. to obtain 1 MΩ with 2000 V max, set 100 kΩ switch to position "10."