The 1692 RLC Digibridge is a benchtop LCR meter for inspecting, evaluating and sorting discrete components at five defined test frequencies. It is particularly suited to common axial and radial leaded parts, which connect through the built-in four-terminal Kelvin fixture so that the meter can measure the DUT rather than rely on a two-wire lead connection.
- Five selectable test frequencies: 100Hz, 120Hz, 1kHz, 10kHz and 100kHz, with .01% frequency accuracy.
- Measures series or parallel R/Q, L/Q, C/D and C/R, allowing the reported value and associated loss quantity to match the required equivalent-circuit model.
- 0.3 V or 1.0 V rms maximum AC test signals, 2 V internal dc bias, and external dc bias of up to 60 V.
- Approximately 2, 4 and 8 measurements/second at SLOW, MEDIUM and FAST rates, with 10-bin sorting for inspection and screening.
- OPEN and SHORT zero calibrations remove fixture, cable and connection residuals at the selected frequency.
It is a fixed-frequency instrument, not a swept-frequency characterization system. Confirm the applicable RLC and DQ accuracy for the selected frequency, rate, circuit mode and DUT value; specifications double below 15 degrees C and above 35 degrees C.
Measure component value and loss at defined test points
The 1692 is intended for fixed-frequency RLC measurement where the test point itself is part of the requirement. Select 100Hz, 120Hz, 1kHz, 10kHz or 100kHz, each with .01% frequency accuracy, and evaluate the component at the frequency relevant to the inspection or acceptance requirement.
Series or parallel R/Q, L/Q, C/D and C/R measurement modes let the engineer choose the equivalent circuit used to report a real, non-ideal component. This matters because the same DUT can produce a different reported primary value when its losses are represented as series rather than parallel loss. The associated Q, D or R result is displayed with the primary R, L or C value, so a component can be checked for both nominal value and loss behavior rather than value alone.
The LED display provides five digits for R, L or C and four digits for D, Q or R. VALUE, ΔRLC and Δ% displays support direct comparison with a stored nominal value; Δ% has a maximum resolution of one part per million. The BIN NO. display shows the assignment resulting from the programmed limits.
Set the electrical conditions for the DUT
The 1692 provides 0.3 V or 1.0 V rms maximum AC test signals. Its CONSTANT VOLTAGE mode uses a 25 ohms source impedance to maintain the AC test level over a wider impedance range. A 2 V internal dc bias is available, and external dc bias of up to 60 V may be applied when the component test requires bias.
Measurements may be continuous, triggered single or averaged, with averaging selectable from 1 to 10 measurements. SLOW, MEDIUM and FAST rates provide approximately 2, 4 and 8 measurements/second. Select rate and averaging according to the required balance between reading rate and the measurement condition: relative to SLOW, the stated RLC and DQ accuracy specifications are multiplied by 2 at MEDIUM and by 5 at FAST. CONSTANT VOLTAGE mode multiplies the basic RLC and DQ accuracies by 2.
Control fixture and lead contributions
The built-in fixture accepts most common axial and radial leaded components and uses four-terminal Kelvin connections to the DUT. This is useful when contact and lead resistance must not be treated as part of a low-value component measurement.
An OPEN circuit zero calibration removes stray capacitance and conductance shunting the internal fixture or an attached fixture or cable. A SHORT circuit zero calibration removes resistance and inductance in series with the test connections. Make new zero calibrations whenever the test frequency or fixture geometry changes, so the correction represents the actual connection arrangement. The 1692 automatically selects the measurement range for best accuracy and resolution; RANGE HOLD is available when a fixed range is required.
The instrument is protected against connection of charged capacitors with up to 1 joule of stored energy.
Turn a measurement into a sorting decision
For incoming inspection or production screening, the 1692 stores a nominal value and bin limits, then assigns the measured result to one of 10 bins. The sorting arrangement comprises 8 pairs of RLC limits, one DQ limit and one fail bin, with GO/NO-GO lights. This allows the primary RLC value to be limited while a D, Q or R criterion checks loss behavior that could be missed by a value-only decision.
Keyboard settings, programmed nominal values and bin limits are retained in non-volatile memory for automatic reentry upon power-up. Keyboard lock prevents inadvertent changes to test conditions and sorting routines. Power-up self-tests verify operation and calibration validity, with coded display errors when a problem is detected.
Integrate with automated test
The optional 1658-9620 IEEE/Handler Interface provides an IEEE-488 bus and handler I/O. All keyboard functions except external-bias control and internal-standard recalibration are programmable from the bus. RLC, DQ and bin data are available in ASCII or Binary format.
For handler-controlled sorting, outputs report bins 0 through 9, ACQ OVER at the end of data acquisition, and EOT when valid bin numbers are available. A handler START input initiates measurement. This gives a handler a defined measurement-complete signal and bin result instead of requiring an operator to interpret each display reading.
The IET difference
The 1692 combines a built-in four-terminal fixture for common axial and radial leaded parts with OPEN and SHORT zero calibrations, selectable equivalent-circuit modes, and programmed bin sorting. These documented functions address the complete fixed-frequency inspection task: establish the required electrical condition, compensate the actual connection arrangement, assess value and loss together, and produce a defined sorting result.
Supplied items and accessories
The 1692 is supplied with a power cord, axial-lead adaptors, bias cable and instruction manual. The built-in test fixture is part of the instrument.
Documented accessories are the 1658-9620 IEEE/Handler Interface, 1689-9601 BNC Adaptor, 1657-9600 extender cable with alligator clips and banana jacks, 1689-9606 extender cable with Kelvin Clips, 1688-9600 extender cable with 874 connectors, 1689-9602 extender cable BNC to ENO, and an SMD Tweezer. The optional 1689-9604 Calibration Kit permits complete recalibration of the internal standards for each measurement range.
For operating details and service information, consult the 1692 RLC Digibridge user and service manual.
Selecting and ordering a 1692
Request a quote or contact technical support with the component type and connection style, required test frequency, expected RLC value and loss parameter, required AC test level and bias, measurement rate, and whether manual binning or IEEE-488/handler integration is required. Also identify the fixture or cable arrangement, because OPEN and SHORT zeroing must represent the actual test connection.
| Specification | Value |
|---|---|
| Measurement parameters and modes | Series or parallel R and Q, L and Q, or C and D or C and R |
| Measurement modes | Continuous, triggered (single), or averaged; choice of 1 to 10 measurements |
| Primary-value display | Five digits for measured R, L or C value; automatically positioned decimal points, units of measurement and minus signs when appropriate |
| Secondary-parameter display | Four digits for D, Q or R; automatically positioned decimal points and minus signs when appropriate |
| ΔRLC display | Difference between measured R, L or C and a user-entered nominal value, with appropriate units; five digits |
| Δ% display | Percentage deviation of measured R, L or C from stored nominal value; five digits; maximum resolution of one part per million |
| BIN NO. display | Single-digit bin assignment number based on measured value and user-entered bin limits |
| Range selection | Automatic for best accuracy and resolution; RANGE HOLD can inhibit automatic ranging |
| Specification | Value |
|---|---|
| Test frequencies | 100Hz, 120Hz, 1kHz, 10kHz and 100kHz |
| Frequency accuracy | .01% |
| AC test signals | 0.3 V or 1.0 V rms maximum |
| CONSTANT VOLTAGE mode source impedance | 25 ohms |
| Internal dc bias | 2 V |
| External dc bias | Up to 60 V may be applied |
| Measurement rates | Approximately 2, 4 and 8 measurements/second for SLOW, MEDIUM and FAST selections |
| RLC and DQ accuracy rate factor | SLOW: 1; MEDIUM: 2; FAST: 5 |
| CONSTANT VOLTAGE accuracy factor | Multiply basic RLC and DQ accuracies by 2 |
| RLC basic accuracy term | 0.1% of reading; stated for SLOW-rate measurements and subject to the documented low-value and high-value added error terms |
| RLC accuracy applicability | Valid when phase angle is low: Q of R < 0.1, Q of L > 10 or D of C < 0.1; assumes OPEN and SHORT zero calibrations at the selected test frequency |
| High D and low Q effect | Multiply RLC accuracy specifications as specified by the instrument documentation, where D and Q are absolute values |
| DQ accuracy at extreme RLC range | When a high-end or low-end RLC error term is not small in comparison with the basic RLC accuracies, multiply D or Q accuracy by RLC accuracy in percent divided by basic RLC accuracy |
| Specification | Value |
|---|---|
| Built-in fixture | Accepts most common axial and radial leaded components |
| DUT connection | Four-terminal (Kelvin) |
| OPEN zero calibration | Removes effects of stray capacitance and conductance shunting the internal test fixture or other fixture or cable connection |
| SHORT zero calibration | Removes effects of resistance and inductance in series in the test connections |
| Charged-capacitor protection | Up to 1 joule of stored energy |
| Sorting | 10 bins: 8 pairs of RLC limits, one DQ limit, one fail bin; GO/NO-GO lights |
| Parameter storage | All keyboard settings and programmed nominal values and bin limits stored in non-volatile memory for automatic reentry upon power-up |
| Keyboard lock | Prevents inadvertent changes to test conditions and sorting routines |
| Self-check diagnostics | Power-up self-tests verify proper operation and validity of calibration; coded error signals on display notify operator if a problem is encountered |
| Optional interface | IEEE/Handler Interface, 1658-9620 |
| IEEE-488 functions | AH1, SH1, T5, L4, SR1, RL2, PP0, DC1, DT1 and C0 |
| IEEE-488 programming | All keyboard functions programmable from bus except external bias control and internal standard recalibration |
| IEEE-488 output data | All RLC, DQ and bin data; ASCII or Binary format |
| Handler outputs | Sorting results, bins 0 through 9; ACQ OVER; EOT; active low open collector drivers rated at 30 V max.; sink 16 mA at 0.4 V; external power and pull-up resistors required |
| Handler input | START, one line; active low: 0V < Vl < 0.4V and +2.5V < Vh < +5V |
| Specification | Value |
|---|---|
| Operating environment | 0 to 50 degrees C, < 85% RH |
| Temperature effect on specifications | Double specifications below 15 and above 35 degrees C |
| Storage environment | -40 to +74 degrees C |
| Altitude | <2000m |
| Installation category | 1 |
| Pollution degree | 1 |
| Power requirement | 90 -250V AC, 50 -60 Hz; automatic voltage switching; 60 watts maximum |
| Net weight | 10 lbs. (4.5 kg) |
| Shipping weight | 15 lbs. (6.8 kg) |
| Supplied accessories | Power cord, axial-lead adaptors, bias cable and instruction manual |
| Available accessories | IEEE/Handler Interface, 1658-9620; BNC Adaptor, 1689-9601; Extender Cable with alligator clips and banana jacks, 1657-9600; Extender Cable with Kelvin Clips, 1689-9606; Extender Cable with 874 connectors, 1688-9600; Extender Cable BNC to ENO, 1689-9602; SMD Tweezer; Calibration Kit, 1689-9604 |